5 Configuration
IM/NAV6S/SS–EN Rev. E
29
Fields
Description
Security type
A page opens with two fields:
Security system
– toggles between 'Basic' and 'Advanced'.
Configuration security
– toggles between 'Password protected' and 'Internal switch
protected'.
Operator level security
Set access to the 'Calibration & Maintenance' and 'Logging' menus.
If set to 'Off', no password is required.
If set to 'On' and 'Security type' is set to 'Basic' an additional 'Operator level password' field is
displayed.
If set to 'On' and 'Security type' is set to 'Advanced', all users are required to enter their user
password to gain access to the 'Calibration & Maintenance' and 'Logging' menus.
Operator level password
Displayed only if 'Security system' is set to 'Basic' and 'Operator level security' is set to 'On'. All
users are required to enter this password to gain access to the 'Calibration & Maintenance' and
'Logging' menus.
The following tabs are displayed only if 'Security System' is set to 'Advanced'.
Reconfigure preset
Passwords are set initially by User 1 (System Administrator) but any user can subsequently change
their own password.
When set to 'Yes' each user must change their password after it is used for the first time following
initial configuration.
Password expiry
Select the number of days that the password is valid for. When a password expires, the user is
prompted to provide a new password.
Inactive user disabling
Select the number of days after which an inactive user's access privileges are de-activated.
Password failure limit
Enter the number of consecutive incorrect password entries allowed by a user. If the number of
incorrect entries exceeds this limit, the user's access privileges are de-activated and can be
reinstated only by the System Administrator (User 1).
Min password length
Sets the minimum length required for user's passwords.