Navigator 600 Silica
Single-stream
Appendix B – Diagnostics and Alarms
68
IM/NAV6S/SS–EN Rev. E
Non-Volatile Memory Error
There is a problem with either the display electronics or the main
board memory.
1.
Power the analyzer down, wait for 10 seconds and power
back up.
2.
If the error persists, contact your local ABB representative.
OFF
ON
Scheduled
zero and
secondary
calibrations
Calibration Factor Too High / Low
A secondary calibration failure has occurred because the calibration
factor is less than or greater than the acceptable limit programmed.
1.
On the 'Calibration' page, check that the secondary
calibration value has been entered correctly.
2.
Remake the secondary standard. Use a higher value solution
if possible, as this is easier to make with less chance of
errors.
3.
Replace the reagents with a fresh batch, especially if this
message appears after the first calibration following the
reagent replacement (applicable to low calibration factor
only).
OFF
ON
Scheduled
secondary
calibrations
Reaction Block Temp Stabilizing
This message is displayed at startup and remains until the reaction
block temperature has stabilized to within 2 ºC (3.6 ºF) of the
programmed control temperature. If the message is still displayed
after 15 minutes:
1.
Check that the sample and ambient temperatures are within
the quoted limits.
2.
Check the 'Heater' value displayed on the 'Measurement
Status' page within 'Diagnostic Information' – see
Section 12.3, page 60.
If the temperature is low, check the heater LED on the main board is
lit. If it is not, contact your local ABB representative.
ON
OFF
Reaction Block Temp. High / Low
>2 °C (3.6 °F) but <20 °C (36 °F) above or below
The measured temperature is more than 2 °C (3.6 °F) but less
than 20 °C (36 °F) above or below the programmed control
temperature.
1.
Check that both the sample and ambient temperature are
within the quoted limits.
2.
If the error persists, contact your local ABB representative.
OFF
OFF
All calibrations
and cleans
Icon
Diagnostic message
Out-of-
Service
Relay
Hold
Relay
Cal. in
Progress
Relay
Cal.
Failed
Relay
Inhibits
Table B.1 Diagnostic Information, Relay Operations and Inhibits (Sheet 3 of 7)