Navigator 600 Silica
Single-stream
5 Configuration
IM/NAV6S/SS–EN Rev. E
33
5.4.2 Streams
5.4.3 Diagnostics
Fields
Description
Stream 1
The page opens with three fields:
Stream 1 Tag – adds an identification for the stream.
Range Low – sets the low range value (0 to 5000 ppb).
Range High – sets the high range value (0 to 5000 ppb).
If the difference between the low and high value is too small, the trace displayed appears very noisy.
The difference between the low and high value must be at least 50 ppb.
Fields
Description
Reagent Bottle Capacity
The
Reagent Bottle Capacity
can be set between 0.1 and 10.0 liters. This should be changed only if
the initial volume of each reagent installed is not 2.5 liters.
This value is used to predict the level of reagent remaining in the bottle.
Reagent Limit
The early warning alarm point for the reagent solutions. Can be set between 1 and 30 days.
Set to a suitable interval to allow new reagents to be ordered.