7.
Enter 10,000ppm into the Total [H
2
O] ppm field. (Figure 92)
8. Enter the values of
δ
D,
δ
18
O and
δ
17
O in PP_MIL (‰) that are labeled on the Water
Standard Bottle. (Figure 92)
9. Click the NEXT button on the lower, right panel of the screen to begin calibration.
(Figure 92)
10. Each step is displayed in the lower-right panel of the calibration screen as the
analyzer performs the calibration. Figure 93 shows the calibration process as a flow
chart.
Figure 93: Calibration Flow
11. When the
Calibration Complete message is displayed, click the CLOSE button.
(Figure 92)
12. Enter
TimeChart by selecting the Display button on the User Interface Control Bar
(Figure 91), and verify that the displayed isotope values correctly correspond to the
values listed on the bottle.
The time of latest calibration is stored in
Reference Gas Settings within
the
Calibration menu for future reference.