USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
106
9. In the
Time between reference cycles pane, use the arrow scroll bar to set the
number of minutes between each reference cycle. (Figure 104)
Figure 104: Time Between Reference Cycles Pane
a. If running the MIU simultaneously with the WVISS, click on the Set by MIU
check box. See the Configuring the WVISS and MIU to run simultaneously
section on page 108 of this appendix for details.
10. Click Save Changes. (Figure 97)
11. Click Close to begin the WVISS configured run. (Figure 97)
a. While the WVISS is operating, the current status of the WVISS is displayed in
the parameter window of the
User Interface Control Bar. (Figure 105)
b. The data is saved to the Data File listed in the parameter window of the
User Interface Control Bar. (Figure 105)
Figure 105: WVISS Status
12. When sampling is complete, disable the WVISS by returning to the
WVISS screen in
the
Setup panel. (Figure 96) Within the WVISS mode pane, use the drop-down
selection box to disable control of the WVISS. (Figure 106)
Figure 106: WVISS Mode Drop Down Menu
13. Click Save Changes. (Figure 97)