Laser Adjust Tab
Use the
Laser Adjust tab to manually adjust the laser’s wavelength to compensate for any
cumulative drift. (Figure 35)
Laser adjustment may be needed for the following reasons:
The laser’s wavelength has drifted beyond the target range of the analyzer.
The analyzer is operated outside the recommended temperature range.
Figure 35 shows the offset between absorption peaks and target lines. For a 2-laser
system, both lasers need adjustment. Figure 35 shows an example for the GLA132-GGA,
which has 2 lasers. Laser A is the top plot and laser B is the bottom plot.
Figure 35: Absorption peaks off of target lines. Laser voltage adjustment needed.