USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
61
Replace the Fuse
If the fuse blows or is otherwise damaged, the analyzer automatically turns off.
To replace the fuse:
1.
Flip the Power switch to OFF on the left side of the analyzer. (Figure 4)
2.
Unplug the 11-30 VDC power connector from the left side of the analyzer. (Figure 4)
3. Unlatch the Case Securing Latches. (Figure 45)
a. Push the latch button.
b. Pull down on the latch.
Figure 45: Case Securing Latches
4. Open the case top. (Figure 46)
Figure 46: Open the Case Top
Содержание GLA132 Series
Страница 1: ... ABB MEASUREMENT ANALYTICS USER MANUAL User Manual ICOS GLA132 Series Ultraportable Analyzers ...
Страница 65: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 65 ...
Страница 99: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 99 ...
Страница 107: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 107 Figure 107 shows the WVISS Disabled Figure 107 WVISS Disabled ...
Страница 130: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 130 Europe United States ...