Appendix I:Isotope Definitions
The CCIA measures the concentration of
12
CO
2
,
13
CO
2
, and OC
18
O. These concentrations are
used to calculate the total CO
2
and the isotope ratios that are reported on the display
screens. The data file output includes the concentrations as well. The terms and their
respective data file name are listed below:
The isotope ratios are reported in ‰ relative to Vienna Pee Dee Belemnite converted to
CO
2
(VPDB-CO2). The standards listed below were taken from IAEA-TECDOC-825.
(R
13
) VPDP-CO
2
0.0112372
(R
18
) VPDP-CO
2
0.002088349077
Total CO
2
is defined as the sum of all the isotopes:
CO
2=
[
16
O
12
C
16
O]+ [
16
O
13
C
16
O]+ [
16
O
12
C
18
O]
The isotope ratios are defined according to:
𝛿
13𝐶
=
(
𝑅
13
)
𝑀𝑒𝑎𝑠
(
𝑅
13
)
𝑉𝑃𝐷𝐵−𝐶𝑂2
−
1 × 1000
𝛿
18𝑂
=
(
𝑅
18
)
𝑀𝑒𝑎𝑠
(
𝑅
18
)
𝑉𝑃𝐷𝐵−𝐶𝑂2
−
1 × 1000
CO
2
X_CO2__ppm
13
C
D13C_VPDB_CO2
18
O
D18O_VPDB_CO2
[
16
O
12
C
16
O]
X_CO2_626__ppm
[
16
O
13
C
16
O]
X_CO2_636__ppm
[
16
O
12
C
18
O]
X_CO2_628__ppm
Содержание GLA132 Series
Страница 1: ... ABB MEASUREMENT ANALYTICS USER MANUAL User Manual ICOS GLA132 Series Ultraportable Analyzers ...
Страница 65: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 65 ...
Страница 99: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 99 ...
Страница 107: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 107 Figure 107 shows the WVISS Disabled Figure 107 WVISS Disabled ...
Страница 130: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 130 Europe United States ...