62
FSM4000
ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
… 8 Operation
… Parameter descriptions
Submenu / parameter
Value range
Remark
Submenu
This menu does not appear with an excitation frequency of 50 Hz
or 60 Hz. Only for SE41F and SE21 transmitters.
Data 70 Hz
Channel
View only
Display of the phase deviation between signal and reference
voltage in µs.
-123 µs
Zero
View only
Display of the adjustment values for zero point, span forward
and span reverse when amplification ‘High’ is selected in the
‘Service / Primary Adjust / Amplification’ menu.
Span >V
Span <R
Zero_2
View only
Display of the adjustment values for zero point, span forward
and span reverse when amplification ‘Middle High’ is selected in
the ‘Service / Primary Adjust / Amplification’ menu.
Span 2 >V
Span 2 <R
Zero_4
View only
Display of the adjustment values for zero point, span forward
and span reverse when amplification ‘Middle Low’ is selected in
the ‘Service / Primary Adjust / Amplification’ menu.
Span 4 >V
Span 4 <R
Zero_8
View only
Display of the adjustment values for zero point, span forward
and span reverse when amplification ‘Low’ is selected in the
‘Service / Primary Adjust / Amplification’ menu.
Span 8 >V
Span 8 <R
Zero_Pre/FIR
View only
Display of the zero point offset with activated filter in ‘Service /
Noise Handling / Prefilter /’ and ‘Service / Noise Handling /
Prefilter / FIR EA’
Zero_Pre
View only
Display of the zero point offset with activated filter in ‘Service /
Noise Handling / Prefilter /’