ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
… 8 Operation
… Data entry
Entering data in short form
Task
Input
Display information
Starting point for ‘Process information’
→
V98.14 %
→
V12.000 m
3
1.
Open menu.
C/CE
The required
arbitrary parameter appears
2.
Search ‘Prog. Level’.
ENTER
‘Prog. level’
Disabled
3.
Switch off ‘Programming protection’,
Select ‘Specialist’.
STEP or DATA
‘Prog. level’
specialists
4.
Set value.
ENTER
‘Prog. level’
specialists
Direct numeric entry
Entry from table
Task
Input
Display information
Task
Input
Display information
5.
Find parameter ‘Qmax’.
STEP or DATA
Qmax
5.
Parameter “Submenu
unit”.
STEP or DATA
Submenu
1800.00 m³/h
Unit
6.
Change parameter
‘Qmax’.
ENTER
Qmax
6.
Change parameter
m³/h.
ENTER
Unit Qmax
m³/h
m³/h
Min 500 m³/h
Max 10000 m³/h
7.
Find desired value, e.g.,
m³/min.
STEP or DATA
Unit Qmax
m³/min
Qmax
6 2 4 0 , 0 0 m³/h
8.
Set value.
ENTER
Unit Qmax
7.
Enter the desired
sequence of digits
6 x DATA
STEP
2 x DATA
STEP
4 x DATA
STEP
STEP
10 x DATA
m³/min
9.
A new value is displayed.
Unit Qmax
m³/min
10.
Exit menu.
C/CE
8.
Set value.
ENTER
Qmax
6240.00 m³/h
9.
A new value is displayed.
Qmax
6240.00 m³/h
10.
Exit menu.
C/CE
11.
Search submenu ‘Prog. Level’.
STEP or DATA
‘Prog. level’
specialists
12.
Power up “Programming protection”.
ENTER
‘Prog. level’
specialists
13.
Select ‘Locked’
C/CE
‘Prog. level’
Disabled