![ABB FSM4000 Series Скачать руководство пользователя страница 75](http://html.mh-extra.com/html/abb/fsm4000-series/fsm4000-series_operating-instruction_4421075.webp)
FSM4000
ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
75
Submenu / parameter
Value range
Remark
Totalizer → F
Numerical entry
0 ... 9999999
Display or input the forward totalizer value.
0.0 kg/h
Overflow ← R
Read only
Overflow totalizer max. 65535. 1 overflow = pulse totalizer
9,999,999 units. Resets the display status and increments the
overflow counter by one.
0
Totalizer ← R
Numerical entry
0 ... 9999999
Display or input the reverse flow totalizer value.
0
Reset totalizer
Select ENTER to reset the forward/reverse flow totalizer and
overflow counters.
Overflow Diff.
Read only
Display the overflow for the diff. totalizer.
0
Totalizer Diff.
Numerical entry
0 ... 9999999
Display or input the diff. totalizer value.
0.0 kg/h
Submenu
This menu does not appear for the PROFIBUS PA or
FOUNDATION fieldbus.
Current output
Current output
0 - 20 mA, 4 - 20 mA
0 - 10 mA, 2 - 10 mA
Select the current output range.
For HART protocol always 4 - 20 mA.
0 - 20 mA
Iout for alarm
Low Alarm
High alarm
Behavior of the current output during an alarm. Low-Alarm is
available only for current output of 2 - 10 mA and 4 - 20 mA. If
error 0 (empty tube) occurs, the current output is set to the value
programmed for the parameter ‘lout during empty tube‘ in the
submenu ‘Detector empty tube‘. If error 3 (flow rate > 103 %)
occurs, ‘High Alarm’ is always signaled.
Low Alarm
Iout low alarm
Numerical entry
Low-Alarm is available only for current output of 2 -10 mA or 4 -
20 mA. Set the current output value during Low Alarm condition.
No selection available if High Alarm is selected for the current
output selection for lout during alarm and if the transmitter
version is standard and the current output range is
0 - 20 mA
0 - 10 mA
.
3.60 mA