Error
messages
D184B111U02 CoriolisMaster
FCM2000
99
Error code and clear
text
Priority Description
Possible Causes
Corrective Measures
Incorrect wiring
(only for remote design)
Check wiring between flowmeter
primary and converter
Error: 7b
T instrument housing
measurement
12 Erroneous
temperature
-
measurement
For the temperature compensation
of the measurement variable Qm
a density of 20°C is used, i.e. for a
fluid temperature near 20 °C the
measurements will be correct
Pt 100 is defective
Check the resistance of PT100 on
the primary flowmeter
Error: 8a
Iout 1 too large
15
The current value is above the
programmed range for current
output 1.
Range setting is too small
Increase range setting
Error: 8b
lout 1 too small
16
The current value is below the
programmed range for current
output 1
Range setting is too small
Increase range setting
Error: 8c
Iout2 too large
17
The current value is above the
programmed range for current
output 2.
Range setting is too small
Increase range setting
Error: 8d
lout 2 too small
18
The current value is below the
programmed range for current
output 2
Range setting is too small
Increase range setting
Error: 9a
Density measurement
9
The measured density of the fluid
in the flowmeter primary is outside
of the specifications
This error usually occurs together
with errors 1 and 9.
See errors 1 and 9
See errors 1 and 9
Error: 9b
Density <0.5 kg/l
10
The density of the medium in the
flowmeter primary is < 0.5 kg/l, the
volume totalizers are stopped
The flowmeter primary is not
completely filled with fluid.
Completely fill the flowmeter
primary
Error: 11a
Sensor A
23
The signal from Sensor A is
missing
Sensor A is defective, or the
amplitude control circuit is open
Measure resistance of Sensor A
For remote design: Check wiring
between flowmeter primary and
converter
Error: 11b
Sensor B
24
The signal from Sensor B is
missing
Sensor B is defective, or the
amplitude control circuit is open
Measure resistance of Sensor B.
For remote design: Check wiring
between flowmeter primary and
converter
Error: 11d
Sensor
5
The signal from at least two
sensors is missing
At least two sensors are defective,
or the amplitude control circuit is
open
Measure resistance of the
sensors.
For remote design: Check wiring
between flowmeter primary and
converter
Error: 12
Concentration
Concentration below 0 % or above
103,125 %.
The measured density or
temperature does not fit in the
given and chosen matrix
Correct the chosen matrix
Error
prot. user trans
Parameters cannot be changed
The operating protection switch is
active
Deactivate the hardware
protection switch.
Pos: 45 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @
Содержание CoriolisMaster FCM2000
Страница 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Страница 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Страница 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Страница 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Страница 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Страница 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Страница 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Страница 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Страница 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...
Страница 123: ...Appendix D184B111U02 CoriolisMaster FCM2000 123 Pos 56 12 Seitenumbruch 0 mod_1126532365768_3101 doc 3830...