Parameterization
64 CoriolisMaster
FCM2000
D184B111U02
Parameter
Value range / input type
Note
DensiMass
code
4289794
If ordered with DensiMass option, this menu
shows the order specific DensiMass code. If the
DensiMass shall be activated after shipping the
meter, please contact your ABB service or sales
organisation.
DensiMass
On
On
Off
Code
invalid
Submenu
Concentration
Medium
Variable
Matrix
Sugar
in
H2O Sodium
hydro.
Alcohol
in
water
Wheat
starch
For the compuation of concentration of 2-phase
liquids a temperature-density-concentration matrix
is required for every blend. ABB has implemented
matrices for some blends already.
Corn starch
Sugar
in
H2O
(BRIX)
Submenu
Variable
Matrix
Submenu
Configuration
Number matrices
1
Selection if 1 or 2 matrices can be entered.
2
2
Number temp.
2 ... 20
2
Definition of number of temperatures (rules for
valid values see 6.4.3 ).
Number conc.
2 ... 20
2
Definition of number of concentrations (rules for
valid values see 6.4.3 ).
Enter conc. in %
Yes
yes
No
In case of a concentration unit separate from %,
additional % concentrations can be selected. This
is mandatory for the calculation of net massflow!
Reset matrix
Resets the matrix values to factory settings.
Содержание CoriolisMaster FCM2000
Страница 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Страница 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Страница 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Страница 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Страница 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Страница 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Страница 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Страница 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Страница 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...
Страница 123: ...Appendix D184B111U02 CoriolisMaster FCM2000 123 Pos 56 12 Seitenumbruch 0 mod_1126532365768_3101 doc 3830...