Parameterization
92 CoriolisMaster
FCM2000
D184B111U02
Calculation of accuracy
The accuracy of the concentration computation depends foremost on the quality of the matrix
data. As density and temperature measurement is the input for the calculation both accuracies
define the accuracy of the concentration measurement.
Example:
Density 0 % alcohol in water (20 °C [68 °F])
998.23 g/l
Density 100 % alcohol in water (20 °C [68 °F])
789.30 g/l
100 %
=
208.93 g/l
0.48 %
=
1 g/l
2.40 %
=
5 g/l
The chosen accuracy class of the density measurement effects the accuracy of the concen-
tration measurement directly.
Example: Input Matrix
10 °C (50 °F)
20 °C (68 °F)
30 °C (86 °F)
0 %
0 °BRIX
0.999 kg / l
0.982 kg / l
0.979 kg / l
10 %
10 °BRIX
1.010 kg / l
0.999 kg / l
0.991 kg / l
40 %
30 °BRIX
1.016 kg / l
1.009 kg / l
0.999 kg / l
80 %
60 °BRIX
1.101 kg / l
1.018 kg / l
1.011 kg / l
The unit of density and temperature corresponds to the chosen units in submenu units.
Pos: 41.13 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @
Содержание CoriolisMaster FCM2000
Страница 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Страница 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Страница 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Страница 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Страница 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Страница 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Страница 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Страница 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Страница 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...
Страница 123: ...Appendix D184B111U02 CoriolisMaster FCM2000 123 Pos 56 12 Seitenumbruch 0 mod_1126532365768_3101 doc 3830...