Installation
D184B111U02 CoriolisMaster
FCM2000
45
Pos: 31.7.5 /Ex-technische Daten/Durchfluss/FCM2000/Hinweise zu Isolation @ 6\mod_1169211927484_3101.doc @ 61924 @ 3
4.6.2 Isolation: MC26.., MC27..
The pipeline and the flowmeter sensor insulation should be installed as shown.
The max. insulation thickness at the flowmeter sensor is 100 mm (4“).
G00331
1
Fig. 44
1 max. 100 mm (4“)
Pos: 31.7.6 /Ex-technische Daten/Durchfluss/FCM2000/FM Installationsklassen @ 5\mod_1167925425812_3101.doc @ 56569 @ 5
FM Installation Class I, Div. 1 / Div. 2
Class I, Div. 1
Area, Zone 1
Class I, Div. 2
Area, Zone 2
General purpose /
Safe Area
G00332
MC27C
MC23O
MC21O
ME21O
ME21T
ME21T
MC21T
MC23T
Abb. 45