Parameterization
78 CoriolisMaster
FCM2000
D184B111U02
Parameter
Value range / input type
Note
Submenu
Current
output
2
In contrast to current output 1, current output 2 is
not HART-enabled and has a fixed current output
range (4 ... 20 mA). In submenu current output 2,
dependent on the measurement variable to be
assigned to the output, only those menus are
displayed which are required for configuring the
output. Current output 2 is always passive.
Not contained in the fieldbus software versions.
Output
of
Qm
Table
Qm
Qv
Density
Temperature
Concentration
Qm Concentration
The selections in this menu are used to define
which of the following variables is to be used for
current output 2.
Qv
→
l = 100 %
0.1 ... 10000000
float
120.00
l/min
In this menu the volume flowrate value is entered
for which the current output is to indicate its
100 % value (20 mA). The menu is only displayed
when the current output is based on the volume
flowrate.
Maximum value: QmMax / minimum density
(0.5 g/cm
3
)
Density
→
l = 0 %
0.5 ... 3.5 g/cm
3
float
0.8 kg/l
In this menu the density value is entered for which
the current output is to indicate its 0 % value
(4 mA). The menu is only displayed when the
current output is based on the density.
Minimum:
0.5 g/cm
3
Maximum: 3.5
g/cm
3
The density value for a current output value of
100 % must be at least 0.01 g/cm
3
larger than the
density value for a current output value of 0 %
Density
→
l = 100 %
0.5 ... 3.5 g/cm
3
float
1.3 kg/l
In this menu the density value is entered for which
the current output is to indicate its 100 % value
(20 mA). The menu is only displayed when the
current output is based on the density.
Minimum:
0.5 g/cm
3
Maximum: 3.5
g/cm
3
The density value for a current output value of
100 % must be at least 0.01 g/cm
3
larger than the
density value for a current output value of 0 %
Temp
→
l = 0 %
-50 ... 180
float
-50.00
°C
In this menu the temperature value is entered for
which the current output is to indicate its 0 %
value (4 mA). The menu is only displayed when
the current output is based on the temperature.
Minimum:
-50 °C
Maximum: 180
°C
The temperature value for a current output value
of 100 % must be at least 10 °C larger than the
temperature value for a current output value of
0 %.
Содержание CoriolisMaster FCM2000
Страница 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Страница 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Страница 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Страница 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Страница 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Страница 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Страница 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Страница 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Страница 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...
Страница 123: ...Appendix D184B111U02 CoriolisMaster FCM2000 123 Pos 56 12 Seitenumbruch 0 mod_1126532365768_3101 doc 3830...