Technical data
104 CoriolisMaster
FCM2000
D184B111U02
Pos: 51 /Überschriften/1/S - U/Technische Daten @ 0\mod_1132904574837_3101.doc @ 3169 @ 11
10 Technical data
Pos: 52 /==== Wechsel ein- auf zweispaltig ==== @ 0\mod_1130421847171_3101.doc @ 3828 @
Change from one to two columns
Pos: 53.1 /Überschriften/1.1/1-spaltig/M - O/Modell FCM2000-MC2 @ 11\mod_1172135774718_3101.doc @ 69568 @ 22
10.1 Model
FCM2000-MC2
Pos: 53.2 /Technische Daten / Datenblatt/Durchfluss/FCM2000/Technische Daten/Messaufnehmer/Modell FCM2000-MC2 @ 11\mod_1167917952234_3101.doc @ 55855 @ 333
G00316
Fig. 63:
Flowmeter sensor for FCM2000-MC2
Meter sizes
"E" (DN 20); "F" (DN 25); "G" (DN 40); "H" (DN 50);
"I" (DN 65); "J" (DN 80); "K" (DN 100); "L" (DN 150)
Measuring ranges for flowrate
Nominal size
Max. measuring range
[Q
max
] in [kg/min]
"E"
DN 20 (3/4“)
0 ...
100
"F"
DN 25 (1“)
0 ...
160
"G"
DN 40 (1 1/2“)
0 ...
475
"H"
DN 50 (2“)
0 ...
920
"I"
DN 65 (2 1/2“)
0 ...
1890
"J"
DN 80 (3“)
0 ...
2460
"K"
DN 100 (4“)
0 ...
4160
"L"
DN 150 (6“)
0 ...
11000
Protection class: IP65/IP67, NEMA 4X
Measurement value deviation for flowrate > DN 20 (3/4“)
(Sizes "F", "G", "H", "I", "J", "K")
± 0.4 % of measured value ± 0.02 % of Q
max
± 0.25 % of measured value ± 0.02 % of Q
max
± 0.15 % of measured value ± 0.01 % of Q
max
± 0.1 % of measured value ± 0.01 % of Q
max
(deviation from measure value + zero error)
Measurement value deviation for flowrate DN 20 (3/4“)
(size "E")
± 0.4 % of measured value ± 0.01 % of Q
max
± 0.25 % of measured value ± 0.01 % of Q
max
± 0.15 % of measured value ± 0.005 % of Q
max
(deviation from measure value + zero error)
Measurement value deviation for flowrate DN 150 (6“)
(size "L")
± 0.4% of measured value ± 0.05 % of Q
max
± 0.25% of measured value ± 0.05 % of Q
max
± 0.15% of measured value ± 0.05 % of Q
max
± 0.1% of measured value ± 0.05 % of Q
max
Operating temperature effect
less than ± 0.006% of Q
max
/ 1 K
Reproducibility of flowrate at rates > 5% of Q
max
0.06 % of measured value for nom. deviation 0.1%
0.10 % of measured value for nom. deviation ± 0.15%
0.15 % of measured value for nom. deviation ± 0.25% and 0.4%
Measuring range density
0.5 ... 3.5 kg/dm
3
Measurement value deviation for density
Standard calibration ± 5 g/l
Expanded density calibration ± 1 g/l
For remote design, the signal cable is already calibrated and
may not be later shortened or lengthened.
This calibration represents a matched pair calibration of transmitter
and sensor.
Reproducibility of density
± 0.1 g/l
Measurement value deviation for temperature
-50 ... 180 °C (-58 ... 356 °F) < 1 °C (33.8 °F)
For ambient temperatures below -20 °C (-4 °F), additional deviation is
to be expected.
10.1.1 Reference conditions
Calibration fluid
Water 25 °C (77 °F) (+ 5 K / - 5 K)
Pressure 0.5 ... 6 bar (7.3 ... 87.0 psi)
Ambient temperature
25 °C (77 °F) (+ 10 K / - 5 K)
Supply power
Line voltage per type plate U
N
± 1 %
Warm-up phase
30 minutes
Installation according to this specification
no visible gas phase
no external mechanical or hydraulic disturbances,
particularly cavitation
Output calibration
Pulse Output
Effect of the analog output on the measurement accuracy
Similar to pulse output ± 0.1% of measured value
Содержание CoriolisMaster FCM2000
Страница 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Страница 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Страница 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Страница 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Страница 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Страница 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Страница 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Страница 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Страница 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...
Страница 123: ...Appendix D184B111U02 CoriolisMaster FCM2000 123 Pos 56 12 Seitenumbruch 0 mod_1126532365768_3101 doc 3830...