C1300
Advanced circular chart recorder
1 Preparation
IM/C1300 Rev. J
5
1.2.1 Recorder Status and Error Page
If any of the start-up self-tests (see Fig. 1.3) fail, the error message 'Fail' is displayed. Refer to Fig. 1.4 for possible cause and remedial
action information.
Fig. 1.4 Start-up Error Messages
Main: CFG Test :Pass
SKey:Type B :Pass
Mod2:Not Fitted
Mod3:Not Fitted
Mod4:Not Fitted
Mod5:Dig Out :Pass
Mod6:RS485 :Pass
Main Board CPU and Configuration Test Results
Software Key (see
Note
below) Identification and Test Result
Module Identification and Test Results
Acknowledging Errors
Mod2:Not Fitted
Mod3:Not Fitted
Mod4:Not Fitted
Mod5:Dig Out :Fail
Mod6:RS485 :Pass
Acknowledge
Input 1
-999.99
deg C
Component
Possible Cause
Action
Module in position 2
Module in position 3
Module in position 4
Module in position 5
Module in position 6
Main Board
Software Key
Analog input and/or analog output calibration is corrupt.
Power down and then up again.
If the fault persists, contact the Company.
Configuration stored in non-volatilememory is corrupt.
Check and correct program data.
If the fault persists, contact the Company.
Note.
The software key is optional hardware that is fitted to enable optional software features – Totalizers, Math
Blocks, System Clock and Timers.
Note.
Acknowledging an Error Message clears the error state but does not rectify the fault. After acknowledging an
error, carry out the appropriate action detailed in the above table.