15
Enter
SET UP INPUT
Control Temp
or
Ion Units
mg/kg
mg/l
ppm
°
C
Enter
CURRENT OUTPUT
or
– –
•
–
Display Zero
– –
•
–
Enter
Display FSD
– –
•
–
Enter
Display as
NH
4
?
N?
NH
3
?
Enter
or
or
6
PROGRAMMING…
6.2.4
Set Up Input Page
Page header.
Advance to next parameter.
Control Temperature
Set the required block control temperature within the range 5 to 45
°
C in 0.1
°
C
increments.
Store.
Advance to next parameter.
Display Type (Ammonium, Ammonia or Nitrogen)
Select display as ammonium (NH
4
+
), ammonia (NH
3
) or nitrogen (N).
Store.
Advance to next parameter.
Ion Units
Select the required display units for ammonia concentration.
Store.
Advance to next parameter.
Display Zero
Set the required value for display within the range 0.05 to 20.00.
Store.
Advance to next parameter.
Display Full Scale
The full scale value is automatically set at two decades above
Display Zero
.
Example–
If the setting is 1.0 the full scale is automatically set to 100.0.
Store.
Return to top of
Set Up Input Page.
or
Advance to
Current Output Page.