PROGRAMMING
6.2.1
Operating Page 1
The values displayed in Operating Page 1 are for viewing only and cannot be altered in this page.
Measurement Units
The measurement units are displayed, e.g. Ammonia ppm.
Advance to next parameter.
Sensor Output
The probe output is displayed in millivolts.
Advance to next parameter.
Control Temperature
The heater block control temperature is displayed in degrees Celsius.
Advance to next parameter.
Slope Check Value
The slope value should be between 80 and 110%. If the value is outside these
limits check the electrode.
Advance to next parameter.
Date
The date is displayed.
Advance to next parameter.
Time
The time is displayed.
Advance to next parameter.
Next Calibration Date
The date when the next automatic calibration is to be carried out is displayed. If the
automatic calibration is disabled, 'OFF' is displayed in place of the date.
Advance to next parameter.
Last Calibration Date
The date of the last automatic or manual calibration is displayed.
Return to top of
Operating Page.
or
Advance to
Security Page
(Section 6.2.3)
.