Maintenance & Service
6-5
2.
Open the door on the right side of the analyzer.
The Wash Head is located at the base of the needle, as shown at
right.
3.
Locate the Wash Head (
Red Circle
, see photo at right).
4.
Visually inspect the Wash Head for any salt or blood build-up
on its bottom surface. For best access, view from below. Refer
to the following figures.
CAUTION:
Do not touch components inside the instrument except as specifi-
cally directed.
Clean Wash Head
No salt or blood build-up
present on Wash Head’s bottom
surface
Wash Head with minor build-up
Wash Head with heavy
build-up