Installing the VetScan HM5
2-9
2.3 Initializing the VetScan HM5
1.
Select
Measure > Run
. A Message that the blank has expired will appear.
2.
Select
OK
to run a blank.
The analyzer will run a blank measurement. When the process is complete, the instrument dis-
plays the result of the blank measurement. (Blank runs are described in more detail in
ning a Blank Measurement” on page 4-5
.)
3.
Select
Accept
if there are no flags and no values highlighted in red, indicating that the analyzer
blank background level is within specifications.
The analyzer then displays a sample measurement screen, as shown, and
is now ready to per-
form an analysis
.
4.
Verify the analyzer’s system settings, and make any needed changes: see
Note:
You will be notified with a pop-up if the blank
falls outside specifications. If this occurs,
refer to
problem persists, call Abaxis Technical Sup-
port.