5-12
Quality Control and Calibration
5.
If QC ranges are for the same control tube, select
Copy
to pull
over the values from the QC menu. If QC values do not match the
lot on the control tube being used for calibration, enter the lot
number, expiration date in MM/DD/YYYY format, and target
values from the HM5 control package insert.
6.
Select
Accept
.
7.
Select
Run Cal
.
When prompted, mix the control, remove the cap, place the tube
in the appropriate tube adapter.
8.
Select
Run
to begin the first calibration run.
When the analysis is complete, the results will appear on-screen.
9.
Select
Run Cal
again. The analyzer will display a reminder to
mix the sample. Mix control well by slowly rolling the tube flat
between your hands 10 times, and then slowly inverting the tube
10-15 times.
10.
Follow the on-screen directions, then select
Run
.
When the second analysis is complete, the results will appear
on-screen.
11.
Select
Run Cal
for the third time. The analyzer will again dis-
play the mix reminder.
12.
Select
Run
.
The analyzer then displays the results of the third calibration.
13.
Select
Calibrate
.
Содержание VetScan HM5
Страница 1: ...Operator s Manual ...
Страница 2: ......
Страница 4: ...1 2 ...
Страница 6: ...1 4 ...
Страница 26: ...2 10 Installing the VetScan HM5 ...
Страница 38: ...3 12 Settings ...
Страница 50: ...4 12 Running a Sample ...
Страница 64: ...5 14 Quality Control and Calibration ...
Страница 80: ...6 16 Maintenance Service ...
Страница 88: ...8 4 Shutdown ...
Страница 112: ...9 24 Troubleshooting ...
Страница 126: ...A 10 Introduction to Veterinary Hematology ...
Страница 136: ...B 10 Operating Principles ...
Страница 154: ...D 16 Veterinary Case Studies ...
Страница 163: ......