Installing the VetScan HM5
2-3
2.1.2
Electrical Requirements
The VetScan HM5 is powered by a standard wall outlet, and requires a power supply of
100–240 VAC, 50–60 Hz, 1.5 A (this is provided with the analyzer).
To avoid power surges and electrical noise, DO NOT plug the analyzer’s power supply
into a circuit that includes a centrifuge or other high-current device.
2.2 Installation
1.
Open the shipping box, and lift the accessory box up and out of the shipping carton.
2.
Remove the stabilizing foam from the top of the HM5.
3.
Use both hands to carefully lift the HM5 out of the foam underneath it and out of the box.
4.
Place the HM5 temporarily on a clean, stable counter.
5.
Remove the plastic wrap from the HM5.
6.
Use the metal latch to open the door on the right side of the HM5.
CAUTION:
Abaxis recommends using the analyzer with surge protection
designed for a computer. In addition, a battery backup is strongly
recommended if the VetScan HM5 will be used in an area prone to
electrical surges or power outages.
Use of an Abaxis provided power supply is required for system
integrity and safety.
CAUTION:
Make sure the analyzer is powered
OFF
before connecting it to the
power supply or any electrical device, such as an external printer,
external keyboard, or computer.
CAUTION:
Place the analyzer in upright position only.
Do not place the unit
on its back, side, or top, or you could cause severe damage.
Содержание VetScan HM5
Страница 1: ...Operator s Manual ...
Страница 2: ......
Страница 4: ...1 2 ...
Страница 6: ...1 4 ...
Страница 26: ...2 10 Installing the VetScan HM5 ...
Страница 38: ...3 12 Settings ...
Страница 50: ...4 12 Running a Sample ...
Страница 64: ...5 14 Quality Control and Calibration ...
Страница 80: ...6 16 Maintenance Service ...
Страница 88: ...8 4 Shutdown ...
Страница 112: ...9 24 Troubleshooting ...
Страница 126: ...A 10 Introduction to Veterinary Hematology ...
Страница 136: ...B 10 Operating Principles ...
Страница 154: ...D 16 Veterinary Case Studies ...
Страница 163: ......