4-4
Running a Sample
4.3 Before Performing an Analysis
4.3.1
Daily Cleaning
1.
Wipe up any spills on the sample rotor.
2.
Keep the instrument and immediate surround-
ings as clean as possible.
4.3.2
Check Tubing
1.
Inspect the reagent tubes and connections to make sure the reagents
can flow freely. Make sure the tubes are not pinched or kinked, or
trapped between or beneath objects.
2.
Check the reagent tubes (except the red waste tubing) for bubbles or
air gaps. A few small, “soda-sized” bubbles are normal, but if any
large bubbles (spanning the width of the tubing) or gaps exist, clear
the affected tubes as follows.
a. Make sure the affected tubes are securely attached at both
ends.
b. From the Home screen, select
Maintenance > Reagent Sta-
tus
.
c. Select
Prime All
to start priming the reagents. Confirmation
will be required. Select
Change Pack
if you are currently
changing the reagent pack, or
Prime
if you only need to clear
bubbles in the tubes. For details on changing the reagent pack,
see “Changing the Reagent Pack” on page 6-9
.
d. Repeat if needed to remove all air gaps and large bubbles from
the reagent tubes.
e. If priming is required more than two times or bubbles persist,
call Abaxis Technical Support.
Note:
Air gaps in the waste tubing (red) are normal.
Sample rotor
Sample tube
adapter
Содержание VetScan HM5
Страница 1: ...Operator s Manual ...
Страница 2: ......
Страница 4: ...1 2 ...
Страница 6: ...1 4 ...
Страница 26: ...2 10 Installing the VetScan HM5 ...
Страница 38: ...3 12 Settings ...
Страница 50: ...4 12 Running a Sample ...
Страница 64: ...5 14 Quality Control and Calibration ...
Страница 80: ...6 16 Maintenance Service ...
Страница 88: ...8 4 Shutdown ...
Страница 112: ...9 24 Troubleshooting ...
Страница 126: ...A 10 Introduction to Veterinary Hematology ...
Страница 136: ...B 10 Operating Principles ...
Страница 154: ...D 16 Veterinary Case Studies ...
Страница 163: ......