![Aanderaa 4330 Скачать руководство пользователя страница 41](http://html.mh-extra.com/html/aanderaa/4330/4330_operating-manual_449151041.webp)
TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
page 41 of 93
The next window shows sensor information like the
Node Description, Product name and number, and
Serial number, refer
Figure 6-7
.
The number of sensor parameters and the processing time can be
viewed in the window.
Select an
Update Interval
for the sensor monitoring.
Press
Start
to start monitoring the sensor readings.
The parameter reading in engineering units is shown as
illustrated in
Figure 6-8
. The reading updates according to the
update interval.
Press
Freeze
to temporarily stop the update; press
Start
to restart
monitoring (Start is the same button as Freeze).
Press
Cancel
to stop monitoring.
The sensor readings should be according to the
description in chapter 6.2 .
Calibration
AADI has discovered that an after-curing occures in the foils during the first 1-3 months after the foil was
manufactured which typically leads to 1-4 % lower readings if the sensors are calibrated before the after-
curing has stopped. As of today methods are in place to cure all foils before they are mounted on the
Optodes and calibrated.
Figure 6-7 Set the Update Interval
Figure 6-8 Monitor Sensor Readings
Содержание 4330
Страница 58: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 58 of 93 Figure A 9 Drawing Cable 4865 ...
Страница 59: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 59 of 93 ...
Страница 62: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 62 of 93 Figure A 13 Drawing Cable 4793 ...
Страница 63: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 63 of 93 Figure A 14 Drawing Cable 5335 ...
Страница 64: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 64 of 93 Figure A 15 Drawing Cable 5336 ...
Страница 65: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 65 of 93 Figure A 16 Drawing Cable 5280 ...