TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
page 30 of 93
5.5 Available Commands for the Oxygen Optodes
Available commands and properties for the Oxygen Optode are given in Table 5-2 and Table 1-2
respectively.
Table 5-2 Main RS232 commands available for the Oxygen Optode.
Command
Description
Do Sample
Execute an oxygen measurement and presents the result
Start
Start a measurement sequence according to current configuration
Stop
Stop a measurement sequence
Do CollectCalDataSat
1)
Collect and save calibration data for 100% saturation
Do CollectCalDataZero
1)
Collect and save calibration data for 0% saturation
Do Calibrate
1)
Execute a two point internal calibration function
Do Test
Internal use
Do AdjustGain
Optimize internal amplification to foil type, only used when changing
between standard foil and fast response foil. Refer chapter 6.1 .
Get
ConfigXML
Outputs info on available properties on XML format
Get
DataXML
Outputs info on available(enabled) parameters on XML format
Get
Property
Output Property value (refer Table 1-2)
Get All
Output all property values
Set
Property(Value,.. Value)
Set Property to Value,… Value
Set
Passkey(Value)
Set passkey to change access level
Save
1)
Store current settings
Load
Reloads previous stored settings
Reset
Resets the sensor with new configuration
Help
Print help information
;
Comment string, following characters are ignored
//
Comment string, following characters are ignored
1) Note that the Save procedure might require up to 20 seconds. Losing power during this period
will cause loss of latest configuration change. Wait for acknowledge, ‘#’, before powering down
the sensor. The save procedure is also executed when running the Do CollectCalDataSat, Do
CollectCalDataZero and Do Calibrate –commands.
Содержание 4330
Страница 58: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 58 of 93 Figure A 9 Drawing Cable 4865 ...
Страница 59: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 59 of 93 ...
Страница 62: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 62 of 93 Figure A 13 Drawing Cable 4793 ...
Страница 63: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 63 of 93 Figure A 14 Drawing Cable 5335 ...
Страница 64: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 64 of 93 Figure A 15 Drawing Cable 5336 ...
Страница 65: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 65 of 93 Figure A 16 Drawing Cable 5280 ...