Sensor Configuration
After installing the sensor turn power on and open the
System Configuration
from the
Menu
button.
Select the newly installed
Oxygen Optode
which should appear in the list of sensors, and tap
Configure
in the lower part of the window.
The
System Configuration
holds a list of output parameters
that can be enabled/disabled by the user, refer Figure 3-2.
Enabled properties (
Yes
) are stored in the data logger:
•
Enable
Air Saturation in engineering units.
•
Enable
Raw data, refer chapter 2.2 .
•
Enable
Temperature in engineering units.
•
Enable HumidityComp
property activates compensation
of vapour pressure in the calculations of the output
parameters. Enable HumidityComp can be set to
No
if
measurements are performed in dry air conditions or if you
like to perform the humidity compensation as a post-
processing operation. Disabling this will typically lead to 1-3
% lower readings depending on the relative humidity of the
air.
The absolute oxygen level,
µ
M, is a default parameter and cannot be disabled, refer TD262a.
To enable/disable a parameter:
Select the output parameter from the list, press
View/Edit
in the lower part of the window, and change the
setting by clicking the box (box is now checked), press
Save
to save and close the window.
We recommend that you enable all parameters in case of later use. The memory card storage capacity is
normally not a limitation for the SEAGUARD
®
. Raw data can be used e.g. to control calibration
coefficients and perform quality control on the data.
3.2 RedReference, Calibration Coefficients and Salinity Compensation
Special property settings and calibration coefficients are found in
User Maintenance
.
Open the
User Maintenance
from the
Menu
button. Select the
Oxygen Optode
from the list of sensors,
and tap
Configure
in the lower part of the window. The user maintenance holds four submenus:
•
Node Description
•
Salinity setting
•
Calibration Coefficients
•
Enable RedReference
Figure 3-2 System Configuration;
Output settings.
Содержание 4330
Страница 58: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 58 of 93 Figure A 9 Drawing Cable 4865 ...
Страница 59: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 59 of 93 ...
Страница 62: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 62 of 93 Figure A 13 Drawing Cable 4793 ...
Страница 63: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 63 of 93 Figure A 14 Drawing Cable 5335 ...
Страница 64: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 64 of 93 Figure A 15 Drawing Cable 5336 ...
Страница 65: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 65 of 93 Figure A 16 Drawing Cable 5280 ...