TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
page 23 of 93
Select the
Node Description
property, press
View/Edit
in
the lower part of the window, and change the setting.
Press
Save
to store the setting when completed.
Node Description
is a user entered text describing the
sensor, placement etc. If using for example a
SEAGUARD
®
sensor string with multiple sensors
connected; renaming the sensor can facilitate analyzing
data. The text is by default set to the product name
followed by product- and serial number, e.g.
Optode
Sensor 4330#52
.
Press
Next>
to continue with the next submenu.
The second submenu in the Oxygen Optode user
maintenance holds a setting for the Salinity:
Select the
Salinity
property, tap
View/Edit
and type the
salinity value. Press
Save
to store the setting.
Note! If you are post compensating for the salinity, you
should set the property value to 0. When sensors are
delivered from the factory they are set to 0 salinity as
factory default.
Press
Next>
to continue with the next submenu.
Figure 3-3 Node Description
Figure 3-4 Set the Salinity
Содержание 4330
Страница 58: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 58 of 93 Figure A 9 Drawing Cable 4865 ...
Страница 59: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 59 of 93 ...
Страница 62: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 62 of 93 Figure A 13 Drawing Cable 4793 ...
Страница 63: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 63 of 93 Figure A 14 Drawing Cable 5335 ...
Страница 64: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 64 of 93 Figure A 15 Drawing Cable 5336 ...
Страница 65: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 65 of 93 Figure A 16 Drawing Cable 5280 ...