TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
page 19 of 93
At normal atmospheric pressure (1013 mbar) the measured
O
2
concentration should not be pressure
compensated. As the sensor is submerged you must perform pressure compensation of 0.0032% per
dbar or for every meter increase of the relative pressure.
The relative pressure = absolute pressure (measured with the optode) – atmospheric pressure (normally
set to 1013 mbar).
Example 1: The measured O
2
concentration with an Optode is 400
µ
M. The measurement was performed
at 1m depth, which is 1dbar relative pressure.
O
2c
= 400×1.000032= 400.012
µ
M
Example 2: The measured O
2
concentration with the Optode is 400
µ
M. The measurement was
performed at 1000m depth, which is 1000dbar relative pressure.
O
2c
=400×1.032= 412.8
µ
M
Содержание 4330
Страница 58: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 58 of 93 Figure A 9 Drawing Cable 4865 ...
Страница 59: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 59 of 93 ...
Страница 62: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 62 of 93 Figure A 13 Drawing Cable 4793 ...
Страница 63: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 63 of 93 Figure A 14 Drawing Cable 5335 ...
Страница 64: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 64 of 93 Figure A 15 Drawing Cable 5336 ...
Страница 65: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 65 of 93 Figure A 16 Drawing Cable 5280 ...