![Aanderaa 4330 Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/aanderaa/4330/4330_operating-manual_449151015.webp)
TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
page 15 of 93
Analog Coef
Float
Coefficients (offset,slope) used for trimming the
analog outputs, default set to 0,1
2)
Enable AirSaturation
BOOL
1
Controls inclusion of air saturation(%) in the output
Enable Rawdata
BOOL
1
Controls inclusion of raw data in the output string
Enable Temperature
BOOL
1
Controls inclusion of Temperature in the output
Enable
HumidityComp
BOOL
1
Enable compensation for vapour pressure,-disable
only for use in dry air or external humidity
compensation
Enable SVUformula
BOOL
1
Refer Appendix 7
Interval
Float
1
Sampling Interval in seconds
DS
No
Owner
String
31
Set the device owner
Location
String
31
Set the location
Geographic Position
String
31
Set the geographic position
Vertical Position
Float
1
User value for describing sensor position
1
) Note! Baud rates lower than 9600 may limit the sampling frequency.
2
) Applicable to 4831 only
1.3 Specifications
For product specifications refer Datasheet
D378 for 4330/4330F, D385 for 4835 and D403 for
4831/4831F on our web site
http://www.aaderaa.com
or contact
You will always find the latest versions of our documentation on the web.
Customers can register to obtain a username and password necessary to gain access to product
manuals, technical notes and software. Please contact
for guidance.
1.4 Manufacturing and Quality Control
Aanderaa Data Instruments products have a record for proven reliability. With over 50 years experience
producing instruments for user in demanding environments around the globe, you can count on our
reputation of delivering the most reliable products available.
We are an ISO 9001 Certified Manufacturer. As a company we are guided by three underlying principles:
quality, service, and commitment. We take these principles seriously, as they form the foundation upon
which we provide lasting value to our customers.
Содержание 4330
Страница 58: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 58 of 93 Figure A 9 Drawing Cable 4865 ...
Страница 59: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 59 of 93 ...
Страница 62: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 62 of 93 Figure A 13 Drawing Cable 4793 ...
Страница 63: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 63 of 93 Figure A 14 Drawing Cable 5335 ...
Страница 64: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 64 of 93 Figure A 15 Drawing Cable 5336 ...
Страница 65: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 65 of 93 Figure A 16 Drawing Cable 5280 ...