TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
page 40 of 93
6.2 Function Test
We recommend that you perform a function test of the sensor operating in air to verify the sensor
readings. Refer chapter 6.2.1 and chapter 6.2.2 , for a description of the function test procedure.
At sea level the oxygen saturation should be approximately 100% in air (90-110 %, depending on air
pressure, local oxygen production/consumption and if the temperature sensor measurements are
representative for the temperature of the foil). The saturation will be significantly lower when you breathe
on or near the sensing foil.
The measured temperature should also agree with the ambient temperature.
6.2.1 SEAGUARD
®
Applications
Leave the Oxygen Optode 4330/4835 mounted onto the SEAGUARD
®
Platform. Power the instrument;
refer TD262a and TD262b for operating instructions.
Open
Administrative Tools-> Sensor Monitor
Sensor Monitor
can be used as a direct reading of the sensor;
the function is mainly used for test purposes.
Select the Optode from the list and press
Start
, refer
Figure 6-6.
Figure 6-6 Select sensors to monitor
Содержание 4330
Страница 58: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 58 of 93 Figure A 9 Drawing Cable 4865 ...
Страница 59: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 59 of 93 ...
Страница 62: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 62 of 93 Figure A 13 Drawing Cable 4793 ...
Страница 63: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 63 of 93 Figure A 14 Drawing Cable 5335 ...
Страница 64: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 64 of 93 Figure A 15 Drawing Cable 5336 ...
Страница 65: ...TD 269 OPERATING MANUAL OXYGEN OPTODE 4330 4831 4835 June 2017 page 65 of 93 Figure A 16 Drawing Cable 5280 ...