19
Weight class and recommended measuring range
When building into a system
When a special weighing pan is to be designed, the weight of the pan and the material to be
weighed should not exceed the weighing capacity of the balance.
To reduce influences of static electricity and magnetism, use materials other than resin and
magnetic material such as iron.
There is a function available to maintain the previous weight value in non-volatile memory,
even if the AC adapter is removed.
When “Zero upon power-on
(
p-tr
)” of “Environment, Display
(
ba5fnc
)” is set to “
1
”, the
previous weight value is displayed upon power-on.
For details, refer to “Zero upon power-on” on page 37.
There is a function available to perform span calibration only, when performing calibration with
a tare on the weighing pan.
When “Span calibration
(
5pn
)” of “Environment, Display
(
ba5fnc
)” is set to “
1
”, span
calibration using the internal mass is performed, with a tare on the weighing pan.
For details, refer to “Span calibration” on page 37.
To set a higher response rate or to batch-weigh small amounts of material, such as a powdery
material, refer to “15. EXTENDED FUNCTION”.
Weight
(Displayed value)
Model
Class
5 kg
2 kg
1 kg
500 g
200 g
100 g
50 g
20 g
10 g
5 g
2 g
1 g
500 mg
200 mg
100 mg
F 1 F 2 M 1 M 2 F 1 F 2 M 1 M 2
M C - 1 0 0 0 M C - 6 1 0 0
F 1 F 2 M 1 M 2
Содержание MC-1000
Страница 1: ...MC 1000 MC 6100 1WMPD4002204B...
Страница 7: ...5...
Страница 68: ...66 20 EXTERNAL DIMENSIONS MC 1000 MC 6100 Unit mm...
Страница 73: ...71 MEMO...
Страница 74: ...72 MEMO...