35
Class
Item and Parameter
Description
0
Not used
at-f
Auto feed
1
Used
Selects whether or not automatic
feed is performed
.
0
No output
1
AD-8121 format
info
GLP output
2
General data format
Selects GLP output method.
For how to set time and date to
be added, refer to the GX series
instruction manual, “10-9 Clock
and Calendar Function”.
0
Not used
dout
Data output
ar-d
Zero after output
1
Used
Adjusts zero automatically after
data is output.
0
600
bps
1
1200
bps
2
2400
bps
3
4800
bps
bp5
Baud rate
4
9600
bps
0
7 bits, even
1
7 bits, odd
btpr
Data bit, parity bit
2
8 bits, none
0
CR LF
Crlf
Terminator
1
CR
CR: ASCII code 0Dh
LF: ASCII code 0Ah
0
A&D standard format
1
DP format
2
KF format
3
MT format
4
NU format
type
Data format
5
CSV format
Refer to the GX series instruction
manual, "10-6 Description of the
Item "Data Format".
0
No limit
t-Up
Timeout
1
1 second
Selects the wait time to receive a
command.
0
No output
erCd
AK, Error code
1
Output
AK: ASCII code 06h
0
Not used
5if
Serial
interface
Ct5
CTS, RTS control
1
Used
Controls CTS and RTS.
Unit
Unit
g
gram
Refer to "5. WEIGHING UNITS".
0
Water temperature
d5 fnc
Density
function
ld in
Liquid density input
1
Liquid density
Available only when density mode is
selected. Refer to the GX series
instruction manual,
"15. DENSITY MEASUREMENT”.
C5 in
Internal mass value correction
Refer to “8. CALIBRATION”.
Displayed only when the internal
mass value correction switch is
set to
1
.
id
ID number setting
Refer to “11. ID NUMBER AND GLP REPORT”.
Factory setting
Содержание MC-1000
Страница 1: ...MC 1000 MC 6100 1WMPD4002204B...
Страница 7: ...5...
Страница 68: ...66 20 EXTERNAL DIMENSIONS MC 1000 MC 6100 Unit mm...
Страница 73: ...71 MEMO...
Страница 74: ...72 MEMO...
Страница 75: ...13...