49
13. CONNECTION TO PERIPHERAL EQUIPMENT
13-1 Connection to the AD-8121B Printer
Set the following parameters to use the AD-8121B printer.
AD-8121B
Class Item
Factory
setting
Mode 1
Mode 2
Mode 3
prt
Data output mode
0
0
,
1
,
2
,
4
,
5
3
0
,
1
,
2
,
4
,
5
ap-p
Auto print polarity
0
ap-b
Auto print difference
0
*
*1
Not
applicable
*1
d-no
Data number output
0 0
0 0
,
1
5-td
Time/date output
0 0
0 0
,
1
,
2
,
3
5-id
ID number output
0 0
0 0
,
1
pU5e
Data output pause
0 0
0 0
,
1
*2
dout
Data output
at-f
Auto feed
0 0
0 0
,
1
bp5
Baud rate
2
2 2 2
ptpr
Data bit, parity bit
0
0 0 0
Crlf
Terminator
0
0 0 0
type
Data format
0
0 0 1
5if
Serial
interface
Ct5
CTS, RTS control
0
0 0 0
*1
Set the parameter when the data output mode is set to the auto print mode (
prt 1
or
prt 2
).
*2
When multiple lines are to be printed, set the parameter to
1
.
AD-8121B settings
Mode AD-8121B
DIP
switches
Description
Mode 1
Prints upon data receipt
Standard mode, statistical calculation mode
Mode 2
Prints using the AD-8121B DATA key or the AD-8121B
built-in timer
Standard mode, interval mode, chart mode
Mode 3
Prints upon data receipt
Dump print mode
DIP switch 3: Handling unstable data
DIP switch 4: Data input specification
ON = To print unstable data
ON = Use the current loop
OFF = Not to print unstable data
OFF = Use the RS-232C
With
data 0
, the weight data can be output to the printer.
With
data 2
, the weight data stored in memory can be output to the printer.
With
data 3
, the calibration report stored in memory can be output to the printer.
Refer to “11. ID NUMBER AND GLP REPORT” for print samples.
Содержание MC-1000
Страница 1: ...MC 1000 MC 6100 1WMPD4002204B...
Страница 7: ...5...
Страница 68: ...66 20 EXTERNAL DIMENSIONS MC 1000 MC 6100 Unit mm...
Страница 73: ...71 MEMO...
Страница 74: ...72 MEMO...
Страница 75: ...13...