19
Data output mode
?
Stream mode
Function setting “F5-0”
The scale outputs the current display data. The data-update rate is approximately 10
times per second. This rate is the same as the display-update.
The scale does not output data while it is in setting mode.
?
Print switch mode
Function setting “F5-2”
When the PRINT switch is pressed while the weighing data is stable (STABLE
indicator is on), the scale transmits the data.
?
Auto-print mode +/- data
Function setting “F5-3“
The scale transmits the weighing data when the display is stable (STABLE indicator
is on) and the data is more than +4d or less than -4d of weight data.
d = minimum weight display (see “12-1 Specifications”)
When in counting mode, “d” is equal to minimum weight display of kg mode.
The next output can be obtained after the display returns to between -4d and +4d.
?
Auto-print mode + data
Function setting “F5-4“
The scale transmits the weighing data when the display is stable (STABLE indicator
is on) and the data is more than +4d of weight data.
d = minimum weight display (see “12-1 Specifications”)
When in counting mode, “d” is equal to minimum weight display of kg mode.
The next output can be obtained after the display returns to below +4d.
?
Command mode
Function setting “F5-1
In the command mode, the scale is controlled by commands that come from the
personal computer and so on.
Command list
“Q” command
Command to request the current weighing data.
Command
Q C
R
L
F
Reply
S T , + 0 0 1 2 3 . 4 5 _ k g C
R
L
F
“Z” command
Same operation as the RE-ZERO or ZERO switch.
Command
Z C
R
L
F
?
This command works as RE-ZERO for the metric models
and as ZERO for U.S.A. models.
“T” command
Same operation as the TARE switch.
Command
T C
R
L
F
?
This command works as TARE for U.S.A. models.
?
The metric models cannot accept this command.
Содержание FG-60KAL
Страница 27: ...25 World map...
Страница 28: ...26 MEMO...
Страница 29: ...27 MEMO...
Страница 30: ...28 MEMO...