10
6. COUNTING MODE
Determines a unit weight (the weight of one piece) from sample pieces, and calculates
how many pieces on the weighing pan using the unit weight.
1. Press the MODE switch to select “pcs”.
(“pcs” = pieces)
2. Press and hold the MODE switch to enter
the sample unit weight storing mode.
3. To select the number of samples, press the
PRINT switch. It may be set to 5, l0, 20, 50
or l00.
4. If necessary, place a container on the weighing
pan, and press the RE-ZERO ( TARE )
switch. Confirm that the right side of the number
of samples shows zero.
5. Place the correct number of samples on the pan
or in the container.
6. Press the MODE switch to calculate and
store the unit weight. Remove the sample. The
scale is set to count objects with this unit weight.
?
The total weight of sample pieces should be more than below regardless
of number of sample pieces.
FG-30K:
25 g
FG-60K:
62.5 g
FG-150K: 125 g
If not, the display shows “
lo ut
” and returns to the previous display.
Increase the number of samples (go to step 3) and try again.
?
If pressing the MODE switch without adding sample pieces or adding
enough weight to be acceptable as a unit weight, the display leaves
counting mode and switches to the next weight unit.
7. Place the objects to be counted on the pan.
?
Unit weigh is maintained even if the scale is powered off.
0
ZERO ?
5 0
ZERO ?
Confirm the display
20 -
ZERO
?
20 0
ZERO ?
20 -
20
234
Содержание FG-60KAL
Страница 27: ...25 World map...
Страница 28: ...26 MEMO...
Страница 29: ...27 MEMO...
Страница 30: ...28 MEMO...