14
9. FUNCTIONS
The scale has Function settings to expand your applications.
The parameters set in the Function settings are maintained even if the power switched
off.
9-1. The procedure for setting parameters
1. Turn the power off.
2. Press and hold the ZERO switch and turn the
power on via the ON/OFF switch. Then, the first
function item and its parameter is displayed.
3. Set the parameter value using the PRINT switch.
?
If you do not need to change the parameter, go to next
step without setting.
4. Press the MODE switch. Then the display goes to
the next function item.
?
In this stage, the new parameter is not stored in
the scale yet.
?
To end changing the parameters, turn the power
off.
5. Repeat the steps 3 and 4 to the last item.
6. After setting the last item, press the MODE switch,
Then,
end
will appear.
7. Press the MODE switch again. The parameters are
stored in the scale, and the scale will automatically boot
up.
?
If you turn the power off before this step is done,
none of the parameters will be changed.
f 1- 1
f 1- 1
Item
Parameter
f 1- 0
f 2- 1
end
888
Содержание FG-60KAL
Страница 27: ...25 World map...
Страница 28: ...26 MEMO...
Страница 29: ...27 MEMO...
Страница 30: ...28 MEMO...