BM series
65
Title block and end block
When the weight values are recorded as GLP data, a "Title block" is inserted at the beginning and
an "End block" is inserted at the end of a group of weight values in the GLP report.
Note If the data memory function is used (except
data
0
), the "Title block" and "End block"
cannot be output.
Key operation
1 With
the
weighing data displayed, press and hold the
key (for approx. 2 seconds) until
5tart
is displayed, and then release the key. The "Title block" is output.
2 The weighing data is output according to the parameter setting of the data output mode (
prt
) of
the function table.
3 Press and hold the
key (for approx. 2 seconds) until
recend
is displayed, and then
release the key. The "End block" is output.
Function
setting
"
info
1
" Function
setting
"
info
2
"
AD-8121 printer format
General format using the balance built-in clock data
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR
Carriage
return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
-----------
A
-
&
-
D
MODEL
-----
BM-300
S/N
-----
01234567
ID
------
LAB-0123
DATE
--
2011/02/20
START
TIME
----
12:34:56
WT
--
+123.4567
--
g
WT
--
+123.4612
--
g
WT
--
+123.4623
--
g
WT
--
+123.4634
--
g
WT
--
+123.4511
--
g
WT
--
+123.4534
--
g
WT
--
+123.4712
--
g
WT
--
+123.4645
--
g
END
TIME
----
12:40
15
SIGNATURE
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
Title block
Manufacturer
Model
Serial number
ID number
Date
Time
Weighing data
Time
Signature
End block