9. Function Switch and Initialization
9.1.
Permit or Inhibit
The balance stores parameters that must not be changed unintentionally.
Example:
Calibration
data for accurate weighing, Data for adapting to the operating environment,
Control data for the RS-232C interface.
The balance is equipped with five switches for the purpose of protecting parameters as shown above.
Each switch can select either "permit" or "inhibit". "Inhibit" protects parameters against
unintentional operations.
In addition to the above, when using balances with software version 1.660 or later, the password
function can be used to restrict calibration functions.
1 Press
the
ON:OFF
key to turn the display off.
2 While pressing and holding the
key and the
RANGE
key,
press the
ON:OFF
key to display
p5
.
3 Press
the
key. Then the balance displays the function switches.
4 Specify the switches using the following keys.
RANGE
key........... To select the blinking digit.
RE-ZERO
key........ To change the parameter of the selected switch.
0
To inhibit changes. (Cannot be used.)
1
To
permit
changes. (Can be used.)
key ............ To store the new parameter and return to the weighing mode.
CAL
key................ To cancel the operation (
Clr
is displayed ). To return to the weighing
mode, press the
CAL
key once again.
(The display shown left indicates the factory settings)
Note 1
: The operation is permitted when logged in as an administrator (
Adm
I
N
).
Function table
0
To inhibit changes to the function table.
1
To
permit
changes to the function table.
Calibration using the internal mass
0
Software ver. 1.00 to 1.64 To inhibit calibration using the internal mass.
Software ver. 1.660 or later
pA55
0
To inhibit calibration using the internal mass.
pA55
1
,
2
To inhibit calibration using the internal mass when
logged in as a user (
U5
E
R
) or a guest (
GUE
5
T
).
1
To permit calibration using the internal mass.
Calibration using an external weight
0
Software ver. 1.00 to 1.64 To inhibit calibration using an external weight.
Software ver. 1.660 or later
pA55
0
To inhibit calibration using an external weight.
pA55
1
,
2
To inhibit calibration using an external weight when
logged in as a user (
U5
E
R
) or a guest (
GUE
5
T
).
1
To
permit
calibration using an external weight.
Automatic self calibration
(for variation of ambient temperature)
0
To inhibit automatic self calibration.
1
To permit automatic self calibration.
Internal mass correction
0
To inhibit correction.
1
To
permit correction.
See Note 1.
See Note 1.