![A&D BM-5 Скачать руководство пользователя страница 42](http://html2.mh-extra.com/html/aandd/bm-5/bm-5_instruction-manual_3099197042.webp)
40
BM
series
8.4.
Calibration Test Using the Internal Mass
This function checks the accuracy of weighing using the internal mass.
Calibration
test
does not perform adjustment.
1 Connect the AC adapter and warm up the balance
for at least one hour with nothing on the weighing
pan.
2 Press and hold the
CAL
key until
CC
in
is
displayed, and then release the key.
3 The balance measures the zero point.
Prevent vibration and drafts to affect the balance.
4 The
measured
zero point data is displayed.
5 The balance measures the internal mass.
Prevent vibration and drafts to affect the balance.
6 The value of the internal mass is displayed.
The normal range of the value is as follows:
Model
Internal mass
Normal range
BM-5
,
BM-5D
5.00000 g
±0.02 mg
BM-20
,
BM-22
20.00000 g
±0.02 mg
BM-252
,
BM-200
BM-300
,
BM-500
200.0000 g
±0.2 mg
7
glp
is displayed and the calibration test report is output
to the RS-232C interface and is stored in memory. Refer to
"GLP output (
info
)" and "Data memory (
data
)" of the
function table.
end
is displayed after the calibration test.
8 The balance automatically returns to the weighing mode.
Press and hold
Release
GLP output
Содержание BM-5
Страница 1: ...Analytical Balance BM 5 BM 5D BM 20 BM 22 BM 252 BM 200 BM 300 BM 500 1WMPD4002195E...
Страница 8: ...6 BM series...
Страница 115: ......