BM series
43
8.7.
Correcting the Internal Mass Value
The balance can correct the internal mass
value within the range shown below. This
function corrects the internal mass value to
conform to an external weight. The corrected
mass value is maintained in non-volatile
memory even if the AC adapter is removed.
The internal mass value is corrected as follows:
Model Correction
reference value Correction range
BM-5
,
BM-5D
5.00000 g
±
1.5 mg
BM-20
,
BM-22
20.00000 g
±
1.5 mg
BM-252
,
BM-200
,
BM-300
,
BM-500
200.0000 g
±
1.5 mg
Example
: With the
BM-300
, 200.0000 g is corrected by +0.6 mg (200.0006 g). When performing a
+0.6 mg correction with a 100 g external weight, the correction value will be +1.2 mg in
accordance with the reference value of 200 g.
1 Perform calibration using the internal mass. Weigh an external
weight and get the correction value.
2 Press
the
ON:OFF
key to turn the display off.
3 While pressing and holding the
key and the
RANGE
key,
press the
ON:OFF
key.
p5
is displayed.
4 Press the
key to display the function switches. Set
the function table switch and internal mass correction switch
to "
1
" as shown to the right using the following keys.
RANGE
key..........To select the blinking digit.
RE-ZERO
key.......To change the value of the blinking digit.
key............To store the new value and return to the weighing mode.
CAL
key...............To cancel the current operation.
5 Press and hold the
RANGE
key (for approx. 2 seconds) to enter
the function table and release the key when
ba5fnc
is displayed.
6 Press
the
RANGE
key several times to display
C5
in
.
7 Press
the
key.
Correct the internal mass value using the following keys.
RE-ZERO
(
+
)key....To increase the value.
MODE
(
-
)key ........To decrease the value.
key...........To store the new value and display the next
menu item of the function table.
CAL
key ............To cancel the current operation and display
the next menu item of the function table.
8 Press
the
CAL
key to return to the weighing mode.
9 Press the
CAL
key to calibrate the balance using the
internal mass.
10 Check that the correction has been performed properly with the
external weight. If the value is incorrect, repeat the correction.
Internal mass correction
Press and hold Press
Press and hold
(approx. 2 seconds)
Weighing
mode
Weighing mode
Function switches
Function table
Corrected value
Standby
Internal mass correction switch
Function table switch
Perform calibration
using the internal mass
200 g
200 g
Correct the internal mass
by +0.6 mg at 200 g.
Calibrate with this
internal mass
The same mass
200.0000
g
200.0006
g
Corrected value