BM series
61
Outputting the clock data of the external device “
info 3
”
When the "GLP output (
info
)" parameter of the function table is set to "
3
", the clock data of
external devices such as a printer or a personal computer can be used instead of the balance
built-in clock data for the GLP/GMP compliant report. This is useful for standardizing the clock
data using the clock function of an external device.
Notes
To output the clock data of an external device, the external device such as an AD-8127
printer or Windows communication tools software WinCT RsCom must have a function to
output the time and date when <ESC>D or <ESC>T is received.
Even with
info 3
set, the data memory function saves the balance built-in clock data for
the calibration report.
Manufacturer
Model
Serial number
Calibration type
Signature
AD-8127 data
Function setting "
info
3
"
General format using the external device clock data
A & D
BM-300
CALIBRATED(INT.)
MODEL
S/N
12345678
ID
LAB-0123
DATE
2018/06/18
TIME
12:34:56
SIGNATURE
Date
ID number
Time