![A&D AD-4532B Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/aandd/ad-4532b/ad-4532b_instruction-manual_3079816064.webp)
62
15-2 AD-4532B-04 RS-232C Serial Interface
The RS-232C serial interface can be connected to a device such as a personal computer, PLC or a
programmable display, to output the values displayed by the AD-4532B and perform controlling,
collection or recoding.
15-2-1 Specifications
Transmission form
Asynchronous, half duplex
Baud rate
1200, 2400, 4800, 9600, 19200, 38400bps
Data bits
7 bits, 8 bits
Parity None,
Even,
Odd
Stop bit
1 bit, 2 bits
Code ASCII
Terminator
CR LF (CR: 0DH, LF: 0AH)
Applicable connector
D-Sub 9-pin
15-2-2 Pin assignment and circuit
12345
Pin No.
Signal
name
Direction Description
2 TXD
Output
Transmit
data
3 RXD
Input
Receive
data
5 SG -
Signal
ground
6
DSR
Output
Data set ready
7 RTS -
8 CTS -
Make a connection
between pin 7 and pin 8.
Others
No
connection
Casing
Shield
External view and pin assignment
15-2-3 Format
Normal W
T
,
±
0
1
2
3
.
4
5
CR
LF
Overflow O L ,
±
9
9
9
9
.
9
9
CR
LF
The value for all the digits when overflowing will be 9.
6 7 8 9
RXD
TXD
DSR
SG
2 3 5
6 7 8
Grounding