![A&D AD-4532B Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/aandd/ad-4532b/ad-4532b_instruction-manual_3079816029.webp)
27
Item and Parameter
Description
0
Not used
COMP ON by control input is not required.
Always COMP ON
f-18
COMP ON
1
Used
Select this when the 2D comparator mode is
used.
0
Upward 2-level judgment
1
Upper/lower limit judgment
f-19
Hysteresis mode
2
Downward 2-level judgment
f-20
Hysteresis time
0
|
99
0
:
Hysteresis mode not used.
1
to
99
: 0.1 second to 9.9 seconds
Comparator
(continued)
f-21
Hysteresis width
0
|
9999
0
:
Hysteresis mode not used.
1
to
9999
: Hysteresis width
f-22
Measured value at
0-V output
-20000
|
20000
Sets the measured value that corresponds to DAV output
value of 0 V.
0
:
Factory setting
f-23
Measured value at
10-V output
-20000
|
20000
Sets the measured value that corresponds to DAV output
value of 10 V.
10000
: Factory setting
0
Outputs DAV value without holding the measured value.
DAV
f-24
DAV mode
1
Outputs DAV value without holding the measured value,
after holding or latching the measured value.
1200
1200 bps
2400
2400 bps
4800
4800 bps
9600
9600 bps
19200
19200 bps
f-25
Baud rate
38400
38400 bps
7
7 bits
f-26
Data bit length
8
8 bits
0
None
1
Odd
f-27
Parity
2
Even
1
1 bit
f-28
Stop bit
2
2 bits
f-29
Reserved internally
Modbus
f-30
Address
0
|
99
0
: No
address
1
to
99
: Address
: Factory settings.