19
7-3 Digital Span Mode
The sensor’s rated output voltage is keyed in. Calibration is performed without using an actual
load.
7-3-1 Setting the minimum division
1. With
d-5p
displayed, press the
↵
key to go to the digital span
mode.
2. When
d 01
is displayed, setting the minimum division is enabled.
Use the following keys to set the minimum division.
1
∧
Increases the value.
1
∨
Decreases the value.
1
↵
Saves the minimum division and goes to step 3.
ESC
Cancels the operation and returns to the
d-5p
display.
7-3-2 Setting the rated capacity
3. Use the following keys to set the rated capacity.
1
>
Selects the digit to set.
1
∧
Increases the value.
1
∨
Decreases the value.
1
↵
Saves the rated capacity and goes to step 4.
ESC
Cancels the operation and returns to the
d-5p
display.
To set the rated capacity
Zero calibration
10000
d-5p
Cap
15000
d-5p
Cap
d-5p
d 01
d-5p
diU
d 02
d-5p
diU