7-2 Calibration Modes
There are three calibration modes.
Digital span mode (
d-5p
)
The sensor’s rated output voltage is keyed in. Calibration is performed without using an actual
load.
Calibration mode (
Cal
)
Zero and span calibration are carried out using an actual load.
Full calibration mode (
fCal
)
Zero and span calibration are carried out using an actual load after minimum division and rated
capacity have been set.
7-2-1 Selecting or changing the calibration mode
1. In the measurement mode, press the FNC key for three seconds or
more to enter the function selection mode.
2. Press the > key to display
d-5p
(digital span mode).
3. With
d-5p
displayed, press the > key to display
Cal
(calibration mode).
4. With
Cal
displayed, press the > key to display
d
fCal
(full
calibration mode).
FNC
4532
d-5p
Cal
fnc
fCal
3 seconds