21
7-4
Calibration Mode
Zero and span calibration are carried out using an actual load.
7-4-1 Zero calibration
1. With
Cal
displayed, press the
↵
key to enter the calibration mode.
2. With nothing placed on the load cell, press the
1
↵
key.
1
↵
Performs zero calibration and goes to step 3.
ESC
Cancels the operation and returns to the
Cal
display.
7-4-2 Span calibration
3. Place a weight on the load cell.
Use the following keys to enter the value of the actual weight used.
Wait for the value to stabilize and press the
1
↵
key.
1
>
Selects the digit to set.
1
∧
Increases the value.
1
∨
Decreases the value.
1
↵
Performs span calibration and goes to step 4.
ESC
Cancels the operation and returns to the
Cal
display.
4. After the span value is saved, the value is displayed for two seconds.
When the calibration is completed,
end
is displayed. Press the
1
↵
key to return to the
Cal
display.
Press
the
1
>
key to go to the full calibration mode.
Press the ESC key to return to the measurement mode.
Cal
Cal0
Cal
------
Cal
Cal0
Span calibration
10000
Cal
5pan
------
Cal
5pan
1.8020
Cal
5pan
end
Cal