![A&D AD-4532B Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/aandd/ad-4532b/ad-4532b_instruction-manual_3079816028.webp)
26
Item and Parameter
Description
0000 0 0 0 0 0
: Permit
1
: Inhibit
Keys
f-12
Inhibit by key
To permit or inhibit zero adjustment by the ZERO key.
To permit or inhibit the hold function by the HOLD key.
To permit or inhibit monitoring the upper/lower limit values.
To permit or inhibit changing the upper/lower limit values.
e.g.:
0101
Inhibits zero adjustment by the ZERO key and inhibits
monitoring the upper/lower limit values.
000 0 0 0 0 0
: Not available
1
: Available
LATCH
f-13
LATCH function
To latch the displayed value.
To latch the comparator output.
To latch the DAV output.
To latch options.
e.g.:
0101
Latches the displayed value and the DAV output while
external LATCH is input.
0
Sample hold
1
Peak hold
2
Bottom hold
f-14
Hold mode
3
Bipolar peak hold
Hold
f-15
Holding time after
releasing hold
0
|
99
0
: No
holding
time
1
to
99
: 0.1 second to 9.9 seconds
0
Comparison when the measured value is stable, excluding the zero
band
1
Comparison when the measured value is stable
2
Continuous comparison, excluding the zero band
Simple
3
Continuous comparison
4
5-level comparison excluding the zero band, by rear panel COMP
1-5 input terminals
5
5-level continuous comparison, by rear panel COMP 1-5 input
terminals
6
5-level comparison excluding the zero band, by time control
f-16
Comparator
mode
2D
7
5-level continuous comparison, by time control
Comparator
f-17
Zero band
0
|
20000
Sets the value for the zero band to be used for comparison.
: Factory settings.