17
7. CALIBRATION
The AD-4532B measures voltage signals from sensors and displays the values. Calibration is
performed on the AD-4532B so that it performs correctly.
7-1 Description of Calibration
Calibration has the following setting items and operations.
Minimum division setting
Selects the minimum division.
Rated capacity setting
Sets the rated capacity.
Zero calibration
Adjusts the indicator so that the measured value will be zero
when no load is applied to the load cell.
Span calibration
Adjusts the indicator so that it can correctly measure the change
in the input voltage generated by loading the load cell.
Notes on calibration
The setting range of output resolution is less than 10000. Even if a value exceeds
10000, it will be displayed.
Perform a periodic check to confirm that the indicator measures correctly. And
perform calibration as necessary.
When the installation site is changed, confirm that the indicator measures correctly,
and perform calibration as necessary.
When performing span calibration using a calibration weight, use a weight having a
mass of more than two-thirds of the maximum measured value to minimize
calibration errors.
Wait for the indicator to stabilize before calibration to ensure correct measurements.
To use the decimal point, use function mode
f-00.